Choosing a Gas Flow Meter: Thermal Mass or Differential Pressure
Which gas flow meter fits our line, and how do we know its readings can be trusted?
Fluid Components International and InTech · Art Womack and Michael Bess
Published Updated
Two technical works from Fluid Components International, a maker of thermal meters, compare differential pressure and thermal dispersion meters for gas. Art Womack puts real world accuracy for a differential pressure meter at plus or minus 5 percent or more, with about 10 to 1 turndown and a higher pressure drop. He credits thermal meters with about 100 to 1 turndown and no added pressure or temperature sensors, and he names limits with wet gas and changing gas composition. Michael Bess adds that a thermal meter is only as trustworthy as its calibration, and that calibrating on air and applying a formula left errors as high as plus or minus 100 percent on other gases in one unnamed maker's meter that he plotted. Both authors work for a company that sells thermal meters, and neither work addresses test stands.
Why a gas reading goes wrong
You add a flow meter to a gas line, and the numbers still do not match what the process uses. The cause is often the choice of meter or the way it was calibrated, not the line.
This article summarizes two works from Fluid Components International, a company that makes thermal dispersion flow meters. Art Womack wrote an undated technical publication that compares differential pressure meters with thermal dispersion meters. Michael Bess wrote an April 2021 InTech article on how thermal mass flowmeters get calibrated. Both authors work for the company, and the company sells the technology they favor. Neither work gives a differential pressure maker's view, so read the claims as one maker's case.
Both works address process gas in plants, such as fuel for heaters, boilers, and cogeneration equipment. Neither addresses a test stand. The questions carry over to a bench, but the figures come from process conditions.
What changes when the fluid is a gas
Womack starts with the fluid. A liquid is close to incompressible, and a uniform liquid holds a steady density. A gas compresses, so its density moves with pressure and temperature. A method that works for liquid does not carry over to gas without thought.
He argues that mass flow is the number worth having for a gas. A volumetric reading will not show how much gas a process used unless density changes are accounted for.
He lists what to weigh when choosing a meter. The list is accuracy, turndown, pressure drop, process temperature, extra sensors, and process connections. He then sets two technologies against it.
Differential pressure meters
A differential pressure meter pairs a transmitter with a primary flow element. The usual element is a sharp edged orifice plate. The square root of the pressure drop across it tracks the volumetric flow rate. Pitot tubes, averaging pitot tubes, v-wedges, and v-cones work on the same idea.
Womack says plants often pick these for gas to keep instruments consistent across a site. That helps maintenance and spares. It does not improve the gas measurement.
To turn volumetric flow into mass flow, you add pressure and temperature transmitters and calculate in the PLC or DCS. That adds sensors, installation work, and cost. A multivariable transmitter can simplify the installation for orifice and averaging pitot elements.
Accuracy needs a close read of the specification. A transmitter may be rated at plus or minus 0.1 percent, but the figure shifts with turndown, the share of flow being measured, long-term drift, temperature effects, and static pressure effects. Womack puts a differential pressure meter at better than plus or minus 1 percent in the best case, and at plus or minus 5 percent or more under real process conditions. He notes that the added pressure and temperature transmitters bring errors of their own. Particles can build up at an orifice edge or block the small ports of a pitot tube. He says a v-wedge handles dirty gas better.
Turndown is about 10 to 1, perhaps 20 to 1. If one process runs high and another low, the meter may measure well only at the high end. A common fix is to stack transmitters of different ranges on one element. That adds cost and complexity.
Pressure drop is lost energy. In a 4 inch line at 3,000 SCFM, Womack puts the loss across an orifice near 50 inches of water column. An averaging pitot tube or v-wedge loses under 20. At low process pressure, these losses can make it hard to hold the minimum flow the system needs.
Heat is the next limit. Most transmitters are rated to 250 degrees F at the measuring cell. Near 500 degrees F you need impulse tubing to shed heat, and he gives a rule of thumb of about a foot of stainless tube per 100 degrees. Hotter service may need a process seal with a non-expanding fill fluid. Either one slows the response, adds cost, and needs correct elevation where condensate forms in the lines.
Calibration checks add hardware. A 5-way manifold between the element and the transmitter lets you isolate the transmitter and apply a known pressure for zero and span checks while the system runs. A 3-way manifold isolates only, so the transmitter must come off to be checked.
Thermal dispersion meters
A thermal dispersion meter uses two precision RTDs. One senses process temperature. The other is heated to a known value above it. With no flow, the gap between them is widest. Flow cools the heated sensor and narrows the gap. The cooling depends on the gas's density, viscosity, thermal conductivity, and specific heat. So the meter reads mass flow directly, with no added instruments or calculations, according to Womack.
He states accuracy as a percent of reading plus a percent of full scale. He says these meters resist long-term drift, are commonly compensated across wide temperature ranges, and show negligible effect from pressure changes. His advice is to run the numbers at your worst process conditions for each technology.
Most applications can expect a turndown of about 100 to 1. One meter then covers the range without stacking. The pressure drop across an orifice can run 5 to 10 times that of a single-point thermal element. In his 3,000 SCFM, 4 inch example, that is about 15 inches of water column for the thermal meter against 50 for an orifice. The gap is widest between 70 and 100 percent of maximum flow.
Standard temperature capability runs to about 350 degrees F. Modified elements reach 500 to 850 degrees F with no added installation steps. Installation is simple. In-line meters come with threaded or flanged ends, and insertion elements mount with a compression fitting. Periodic calibration is not required, though makers may recommend a check every 12 to 18 months. For continuous or dirty processes, a packing gland and ball valve assembly lets you pull an insertion element for calibration or cleaning.
The limits are real. These meters do not suit liquids, slurries, or saturated steam. They fit dry gas, gas with little moisture, and superheated steam. They are calibrated for a stated gas composition, such as air, hydrogen, oxygen, or natural gas as methane and ethane. If the composition changes, readings stay repeatable but lose accuracy. A correction factor may bring accuracy back within acceptable limits for some processes. Condensate in the line can cool the heated sensor and give false readings. Sensor placement can reduce the effect, and some cases need knock-out pots or filters.
Piping affects both
Womack says the straight run before and after the meter matters for any technology. For lines up to 6 inches, a run of 20 diameters upstream and 10 downstream is normally accepted. Larger lines call for 15 upstream and 7.5 downstream. A 7.5 foot run of 3 inch pipe is realistic. A 22.5 foot run in 12 inch pipe is harder to find. Elbows and valves distort the velocity profile and reduce accuracy. Elbows out of plane also add swirl.
Flow conditioners can make up for a short run. The common types are perforated plates, screens, vanes, tube bundles, and tabs. Plates and screens create a measurable profile but handle swirl poorly. Tube bundles and vanes handle swirl better but pass on part of a distorted profile. Womack says only the tab design has been shown to remove both effects. Conditioners add pressure loss, and the loss is higher with perforated plates than with tabs. Some thermal meter makers build conditioning into the sensing element. That cuts the straight run needed and gives a meter calibrated for your conditions.
Calibration decides how far to trust a thermal meter
Bess starts from the same physics. A thermal mass flowmeter reads the cooling of a heated element by moving gas, and that cooling depends on the gas's thermal conductivity, specific heat, density, and viscosity. Sensors also differ from unit to unit. He compares them to fingerprints. Even with tight tolerances and automated assembly, the heat transfer paths vary, so one standard gas correction factor cannot cover them.
An actual gas calibration needs a laboratory with traceable flow stands. That is costly, more so for hazardous or flammable gases, and the gas and the energy to flow it at set temperature and pressure add recurring cost. Many makers skip it and run an equivalency calibration. They flow air at ambient conditions and apply a formula to set the response for other gases.
Bess says that method infers the cooling effect and does not reproduce the real heat loss in the actual gas. Corrections for pressure and temperature extremes add more uncertainty. He cites ISO 14511, Section 8, as stating that best practice is an actual gas calibration at actual process conditions when feasible.
He does not recommend equivalency calibration where conditions are somewhat unstable, where flow velocity profiles sit in the transitional range, or where the calibration gas and the service gas relate nonlinearly. Turndowns above 10 to 1 reach past a simple linear correction, and a single factor correction fails there.
His example is a plot of accuracy for a meter from a global maker he does not name, which lets users pick a gas from a menu. Air and nitrogen read near zero error because the base calibration was done in air. With other gases selected, he says the error after the formula was applied can be as high as plus or minus 100 percent. For some gases at slightly elevated temperatures, he says the nonlinearity range can vary by as much as 30 percent, so a single correction factor would not hold across the flow range. He then points to a plot of an FCI Model ST100 calibrated on actual natural gas and calls the improvement exceptional. The comparison is between another maker's meter and FCI's own. The plots are images, and this summary relies on Bess's text and does not reproduce them.
For critical applications, Bess says a buyer has a right to ask how the meters will be calibrated. The questions are on what traceable equipment, by what methods and conditions, and to which standards. He suggests touring the calibration laboratory and meeting the engineers, and having a factory representative review the application and the meter location.
What to take from it
- Decide whether you need mass flow
- For a gas, mass flow tracks the amount used. A volumetric element needs pressure and temperature inputs to get there.
- Test accuracy at worst case conditions
- Womack puts a differential pressure meter at better than plus or minus 1 percent in the best case and plus or minus 5 percent or more in real service.
- Match turndown to your flow range
- Check whether you would need stacked transmitters. Womack gives about 10 to 1 for a differential pressure meter and about 100 to 1 for a thermal meter.
- Count pressure drop as a cost
- It wastes energy and can limit minimum flow at low process pressure.
- Check the gas before choosing a thermal meter
- Ask about moisture and condensate. Ask what composition the meter was calibrated for and whether it will change.
- Check the straight run first
- Plan flow conditioning if the run is short, and count the pressure loss the conditioner adds.
- Ask how a thermal meter was calibrated
- Ask whether it was an actual gas calibration or an equivalency calibration. Ask on which gas and across what range.
- Read both works with their source in mind
- Both come from a thermal meter maker. Womack advises bringing one or two experienced manufacturers into the first evaluation, and that advice comes from a manufacturer.
References
- Womack, Art. Flow Meter Selection for Improved Gas Flow Measurements: Comparison of Differential Pressure and Thermal Dispersion Technologies. Fluid Components International. Undated. https://www.fluidcomponents.com/assets/media/Technical-Articles/Flow-Meter-Selection-for-Improved-Gas-Flow.pdf
- Bess, Michael. Calibrating Thermal Mass Flowmeters. InTech, April 2021. https://www.isa.org/intech-home/2021/april-2021/departments/calibrating-thermal-mass-flowmeters