How to Select a Flow Meter: Five Questions to Answer Before You Compare Models
Which flow meter fits this application, and what do we need to know before we choose?
DDM Sensors · DDM Sensors Team
Published Updated
A flow meter should be chosen from the application, not from the measuring principle. Decide first whether volume or mass flow is needed, since gases with changing pressure or temperature point to mass flow. Then work through five criteria from the DDM Sensors guide: measuring range, medium, installation, accuracy, and standards. Check the straight pipe run and the basis of the accuracy figure early, because both can rule out a meter that otherwise fits.
Where flow meter selection goes wrong
A flow meter catalog lists a dozen measuring principles. It is easy to start with the principle you know best and work backward to the application. That order invites a meter that is wrong for the medium, too wide in range, or fitted where the pipe run is too short.
DDM Sensors, a German maker of pressure and flow sensors, publishes a guide that works the other way. It starts from the application and moves through five selection criteria: measuring range, medium, application, accuracy, and standards. This article follows that guide and adds one earlier decision, which is whether to measure volume or mass.
Decide first: volume flow or mass flow
Volume flow is the volume passing per unit of time, such as l/min or m³/h. The guide says it fits best where density is roughly constant, as with water.
Mass flow is the mass passing per unit of time, such as kg/h. The guide says it is the better choice when density changes with pressure or temperature, which is common with gases. For a gas whose density moves with pressure or temperature, it advises measuring mass flow.
The guide groups common meters this way.
- Volume flow meters
- Impeller and turbine meters turn flow into rotation, and the speed is proportional to flow rate. Oval wheel meters pass a known volume per rotation. Ultrasonic meters use the transit time difference between signals sent with and against the flow. Electromagnetic meters use Faraday's law, so the liquid must conduct.
- Differential pressure and vortex meters
- Differential pressure meters create a pressure drop across an orifice, Venturi or nozzle. The guide calls the principle robust and suited to high flow rates. Vortex meters count the vortices shed behind a bluff body, and the shedding frequency is proportional to flow velocity.
- Mass flow meters
- Coriolis meters vibrate a tube and read the phase shift that flow causes. The guide says they are independent of pressure, temperature and density changes, and suit liquids, gases and steam. Thermal meters measure the heat carried away from a heated element by a gas.
Question 1: what is the measuring range?
The measuring range is the span of flow in which the meter holds its stated accuracy. The guide warns about both ends. A range that is too small may not reach the maximum flow, or may overload the meter. A range that is too large can cost accuracy at the low end.
The ratio between the maximum and minimum flow the meter measures precisely is the turndown ratio. The guide's example is 10:1, which means reliable readings between 1 and 10 l/min. A high turndown ratio helps where flow varies a lot.
The practical step is to find the expected minimum and maximum flow of the application first. Then pick a meter whose range and turndown cover both.
Question 2: what is the medium?
The medium often decides the principle. The guide lists these pairings.
- Low viscosity liquids such as water or fuel
- Electromagnetic, ultrasonic or turbine meters. Impeller meters suit neutral, slightly aggressive liquids with few solids.
- Highly viscous liquids such as oils or adhesives
- Positive displacement meters or Coriolis meters. Oval gear meters also suit adhesives and oils.
- Gases such as air or natural gas
- Thermal mass, vortex or differential pressure meters.
- Steam
- Vortex or differential pressure methods, because pressure and temperature can vary widely.
- Multiphase media such as emulsions and slurries
- Specialized Coriolis or ultrasonic meters with multiphase compensation.
Other properties narrow the list further.
- Electrical conductivity
- Electromagnetic meters need a minimum conductivity, and changes in conductivity can cause errors. Ultrasonic meters are the alternative for non-conductive liquids.
- Solids and bubbles
- Particles can impair some methods. The guide points to displacement meters or special Coriolis or ultrasonic meters. It says to avoid air bubbles with oval gear meters.
- Chemical composition
- Wetted materials must resist the medium to avoid corrosion or other damage.
Question 3: where will it be installed?
The guide treats the installation as a selection criterion in its own right.
- Space and layout
- The meter must fit the available space. Compact versions join the sensor and transmitter. Separate versions let you mount the electronics elsewhere.
- Temperature
- The meter must suit both the medium temperature and the surroundings.
- Connection
- Threaded, flanged or clamp connections must match the existing pipe.
- Hazardous or hygienic areas
- A potentially explosive area calls for an ATEX-certified meter. The guide says hygienic work in pharmaceutical and food settings suits non-contact methods.
- Straight pipe
- Oval gear meters need no inlet or outlet section. Impeller meters need minimum straight lengths before and after the measuring point.
Question 4: how accurate must it be?
The guide ties required accuracy to the job. Billing, and dosing of expensive or critical substances, need high accuracy. Simple monitoring may not. Higher accuracy usually costs more, so the guide recommends a compromise that fits the application.
It names five figures to read in a data sheet.
- Measurement deviation
- The maximum difference between measured and actual flow. It is given as a percentage of the measured value or of the full scale value, so check which one a data sheet uses.
- Repeatability
- How consistently the meter returns the same value under the same conditions.
- Linearity
- How closely the readings follow an ideal line across the range.
- Hysteresis
- The difference between readings with rising and falling flow.
- Long-term stability
- The drift in accuracy over time.
Question 5: which standards apply?
Standards govern safety, compatibility and the comparability of results. The guide names several.
- IP protection classes, such as IP67, for dust and water ingress.
- The ATEX Directive 2014/34/EU, for devices in potentially explosive atmospheres. It classifies them by zone and requires markings that state the degree of protection.
- DIN EN ISO 5167, for differential pressure measurement.
- OIML R117 and MID 2014/32/EU, for accuracy in trade and for devices subject to mandatory calibration.
The guide advises finding out which ones apply before choosing a meter.
Where errors come from after selection
A well chosen meter can still read wrong. The guide lists the common causes.
- Calibration that is out of date or inaccurate.
- Installation with the wrong position, or with inlet and outlet sections that are too short.
- Temperature and pressure swings, which distort volumetric meters on gas.
- Bubbles, particles or deposits in the medium.
- Vibration, which affects Coriolis and electromagnetic meters in particular, and electromagnetic fields.
- Sensor ageing and signal interference.
The guide's remedy is the same list in reverse. Choose the right method, install it carefully, calibrate regularly, and maintain it.
Take-homes for a small team
- Settle volume or mass first
- Gas with changing pressure or temperature points to mass flow.
- Write the application down before comparing models
- Minimum and maximum flow, medium properties, temperature, connection, hazardous area, required accuracy, and applicable standards.
- Check the straight pipe run early
- It can rule out a meter that otherwise fits.
- Read the accuracy figure with its basis
- Percent of reading and percent of full scale are not the same.
- Pay for the accuracy the job needs
- Dosing and billing justify it. Basic monitoring may not.