Differential vs Gauge Pressure Sensing in a CPAP Device
Why does a CPAP device need both differential and gauge pressure measurements, and what changes when one sensor does both?
Superior Sensor Technology · Superior Sensor Technology Team
Published
A PAP device needs a mask pressure reading and a breath flow reading. The Superior Sensor Technology CP Series puts a gauge channel and a differential channel in one package, each with four programmable ranges. The manufacturer's datasheets give looser error and drift figures than its product page. Build the first error budget from the datasheet, and confirm with the manufacturer which figures govern your part.
Two questions a PAP controller answers every cycle
A positive airway pressure device has to know two things at once. It has to know whether the mask holds the pressure the therapy calls for. It also has to know whether the patient is breathing in or out, and how hard. The first answer is a static pressure referenced to the room. The second is a small pressure difference across the flow path.
Most designs use two sensors for this. Superior Sensor Technology, a pressure sensor maker in Los Gatos, California, puts both measurements in one package in its CP Series. This article summarizes what the company's product page and two datasheets say. It then organizes the numbers for an engineer deciding whether a dual-channel part fits a PAP design. The specifications are the manufacturer's. Check them against your own validation data.
Why the device needs both measurements
Superior's product page explains the split this way. Differential pressure, measured across the flow path, carries the dynamic signal of breathing. That includes inspiration, expiration, flow rate, and the pattern of airway obstruction. Gauge pressure, referenced to ambient, carries the static pressure at the mask. Without the differential channel, the device cannot detect breath events or compute flow. Without the gauge channel, it cannot confirm that the target pressure is present at the mask.
The CP Series holds two independent channels in one package. Per the datasheets, the differential channel offers four programmable full-scale ranges from ±250 Pa to ±2.5 kPa. The gauge channel offers four from ±2 kPa to ±6 kPa. A mode register on the sensor selects the range for each channel. The differential options are 250 Pa, 500 Pa, 1.0 kPa, and 2.5 kPa. The gauge options are 2.0, 3.0, 4.0, and 6.0 kPa.
Ranges, filters, and timing
Each channel also has four bandwidth filter corners: 25, 50, 100, and 250 Hz. The product page counts 64 configurations per device. That is four differential ranges, four gauge ranges, and four filter options. The datasheet mode register carries separate filter bits for each channel, so the count depends on how you tally. Either way, ranges and filters are set by software command. One hardware part can serve different therapy pressure windows.
Range selection matters because accuracy is quoted as a percentage of the selected range. The datasheet gives an example. Select the 250 Pa range on the differential channel, and the span is 500 Pa, from -250 Pa to +250 Pa. Typical accuracy of 0.05% of that span is 0.25 Pa. A tighter range gives a smaller absolute error, so match the range to the signal you need to resolve.
Three timing details come from the datasheets.
- The internal update rate is fixed at 498 Hz, about 2 ms. It does not change with range or filter settings. The product page rounds this to a 2 ms response.
- The datasheet suggests sampling at least twice the selected filter corner frequency.
- A mode change has one cycle of latency. The data from the cycle right after the write does not reflect the change. The second cycle does.
The output is a 16-bit signed integer. The selected full scale sits inside the middle 90% of the code range. The datasheet gives the conversion as counts divided by (0.9 x 2^15), multiplied by the selected range. Its example: 3,647 counts on a 500 Pa range is 61.832 Pa.
Reading the accuracy and drift numbers
The table shows datasheet values for the CP201 and CP301 (DS-0005F) at 3.3 V and 25 C. The CP203 and CP303 datasheet (DS-0012A) lists the same figures.
Parameter Differential, typ / max Gauge, typ / max Accuracy, % of selected range 0.05 / 0.1 0.05 / 0.1 Total error band, % of full span 0.1 / 0.3 0.15 / 0.5 Long-term stability, first year, % of full span 0.1 / 0.25 0.15 / 0.4 Thermal hysteresis, % of full span, typ 0.05 0.15 Position sensitivity, Pa, typ 2.0 2.0The datasheet defines "full span" as the whole span of the channel. Its example uses 5,000 Pa for the differential channel, which is ±2,500 Pa. Applying the same convention to the gauge channel's ±6 kPa gives 12,000 Pa. On that basis, the typical total error band works out to about 5 Pa for the differential channel and about 18 Pa for the gauge channel. This is arithmetic on the datasheet's figures, not a number Superior publishes.
The product page quotes tighter figures. It lists a total error band of 4 Pa differential and 10 Pa gauge. It lists long-term stability of 2 Pa and 12 Pa after the first 12 months. It also lists a short-term error band of 1.25 Pa and 6 Pa, measured over 24 hours after auto-zero. The page does not say how it reaches these values, and they do not match the datasheet. Use the datasheet numbers for a first error budget. Ask the manufacturer which set governs your part and test conditions before you rely on the tighter ones.
Drift deserves attention on the gauge channel, because that channel verifies delivered therapy pressure. A typical therapy pressure of 10 cmH2O is about 980 Pa. The page's 12 Pa gauge drift is about 1.2% of that. The datasheet's typical first-year figure, about 18 Pa by the arithmetic above, is about 1.8%. The datasheet also says drift is largest in the first year and lessens after. By year 10, total typical drift is twice the first-year value. The page says this stability supports a full annual service interval without recalibration. That is the manufacturer's claim. Compare it with your own therapy pressure tolerance.
On resolution, the datasheet table gives 16 bits with a typical effective number of bits of 15.7 at a 25 Hz filter. The front pages of the two datasheets quote different effective resolution figures, 17.5 bits and 16.3 bits. Use the table value.
Interface and integration details
- Interface selection. The sensor reads the nSS pin after power-on reset. High selects SPI. Low selects I2C, and grounding nSS is an accepted way to do that.
- Startup. No communication should take place during the interface detect delay, up to 40 ms. The first reading is settled up to 60 ms after reset.
- SPI reads. Each read sends the mode and rate bytes. The first two bytes returned are the differential pressure. Continuing the transfer returns the gauge pressure in bytes 3 and 4, then the mode byte, temperature, model, serial number, and build number. Only the first two bytes sent to the sensor are used.
- I2C addressing. The default address is 0x28 with the SCK line grounded. A resistor divider on SCK selects nine other addresses from 0x29 to 0x31.
- Data available pin. A rate control register sets the divisor of the 498 Hz internal rate for the DAV pin.
- Supply. The suggested operating range is 2.8 to 3.5 V. Typical supply current is 6.5 mA. The datasheet recommends a 0.1 uF decoupling capacitor near the supply pads.
- Temperature. Compensation covers 5 to 50 C. The operating range is -20 to 85 C.
- Pressure limits. The differential channel has a proof pressure of 10 kPa and a burst pressure of 30 kPa. The gauge channel has 250 kPa and 750 kPa.
- Media and assembly. The part measures dry air and non-aggressive gas. Wetted materials include epoxy, nylon, RTV, and silicon, with gold and aluminum on some ports. Reflow peak temperature is 255 C, moisture sensitivity level is 3, and ESD rating is 2 kV human body model.
Housings and models
The CP Series comes in two housings. The SM04 has three ports, and one port is shared between the differential and gauge channels. The page says this suits circuits where both measurements reference a common point. The SM06 has four ports with dedicated ports for each channel. It suits designs where the flow tap and the mask pressure tap sit at different points in the circuit. In both, a bottom port provides the ambient reference for the gauge channel.
The models differ in added features.
- CP201 and CP301 are the baseline parts. They leave pressure control and breath detection to system software.
- CP203 and CP303 add snore detection. The datasheet shows a one-byte snore output, an unsigned energy magnitude, so the host still decides what magnitude counts as snoring. The source can be the gauge or differential channel. There are eight band pass options, from 10 to 250 Hz up to 150 to 250 Hz. There are four low pass corners from 1 to 5 Hz, a gain setting, and an optional peak detector. The product page describes the output as a flag from the differential channel. The datasheet is more detailed, so this article follows it.
- CP202 and CP302 add closed-loop motor control and an overpressure indicator, according to the page. The page says they are offered through a collaborative development program for high-volume production. No datasheet for them was part of this review.
The EB02 evaluation board supports the CP201, CP203, CP301, and CP303.
Take-homes for a design review
- Choose the housing from your pneumatic layout first. If the flow tap and mask tap are separate points, that points to four ports.
- Select the smallest differential range that covers your expected flow signal. Accuracy is a percentage of the selected range.
- Build the first error budget from the datasheet figures. Ask the manufacturer about the page's tighter numbers before using them.
- Compare gauge channel drift against your therapy pressure tolerance, not against the percentage of full span.
- Check your circuit's pressure transients against the 10 kPa differential proof pressure.
- Plan firmware for the one-cycle mode latency and the 60 ms settling time. Sample at least twice the filter corner.
- If you need snore detection, work from DS-0012A. It documents the snore registers. DS-0005F does not.
References
- Superior Sensor Technology. CP Series CPAP Pressure Sensors, product page. https://superiorsensors.com/pressure-sensors/cpap/
- Superior Sensor Technology. DS-0005F, CP Series Dual Low Pressure Sensor data sheet for CP201 and CP301, Revision F. https://superiorsensors.com/wp-content/uploads/DS-0005F.CP-Series.pdf
- Superior Sensor Technology. DS-0012A, CP203/CP303 CP Series with Integrated Snore Detection data sheet, Revision A. https://superiorsensors.com/wp-content/uploads/DS-0012A.CPx03-Series.pdf