3D Printed Silicone Parts: What One Data Sheet Does and Does Not Tell You
We need soft silicone parts without molding tooling. What does a data sheet for a 3D printed silicone tell us, and what does it leave out?
Published Updated
Stratasys publishes mechanical, biocompatibility, and aging data for a printable silicone. The sheet is a vendor document and leaves gaps. Its biocompatibility results name steam-sterilized test articles, its aging chart has no data labels, and some test methods need confirming. Treat it as a starting point, and ask Stratasys for the methods, the test article, and the aging data before designing around it.
A soft silicone part, without molding tooling
A team needs a short run of soft parts. A seal, a gasket, a watch band, a cover that touches skin. Silicone is the usual material, and molding it means tooling. For a few hundred parts, or a design that is still changing, tooling is a hard cost to accept.
Printing offers a way around it. Stratasys, the company behind the Origin P3 DLP printer, publishes a data sheet for a silicone made for that printer. The material is P3 MED Silicone 25A, which the sheet credits to Shin-Etsu. Stratasys says it suits the gap between prototyping and low-to-mid volume manufacturing, without tooling.
This article reads that data sheet for an engineer deciding whether to design around it. It reports what Stratasys states, flags what the sheet leaves open, and lists what to ask before the material goes into a part. Every figure comes from Stratasys. The sheet is a vendor document and offers no independent testing.
What the sheet describes
Stratasys describes a white, single-component UV resin that needs no mixing. The sheet reports curing shrinkage under 1 percent, no odor, and low tackiness. The material was developed exclusively for Origin, so it runs on that system and on no other printer the sheet names.
Stratasys calls the material real additive silicone with properties similar to molding silicones. The sheet does not back that statement with a side-by-side comparison against any molded grade. One detail is also worth noting. The material name ends in 25A, and the sheet reports a Shore A hardness of 23. The sheet does not explain the difference.
The listed uses start with skin-contact medical devices, such as hearing and audiology devices, bolus, respiratory products, and orthotics and prosthetics. The sheet also lists anatomical models, medical device prototyping, wearables, smartwatch bands, earphones, molds, small-batch production, and seals and gaskets for medical and consumer electronics.
Before printing, the liquid resin has a viscosity of 13,000 at 23 °C and a density of 1.3 g/cm³. The sheet prints the viscosity unit as mP·s, which is not a standard viscosity unit. Confirm the unit with Stratasys before relying on the figure.
The mechanical values
Stratasys reports these values, each with the method it lists.
- Tear resistance is 14.8 kN/m. The method listed is ASTM D638.
- Ultimate tensile strength is 4.4 MPa. The method listed is ASTM D638.
- Tensile elongation at break is 835 percent. The method listed is ASTM D638.
- Shore A hardness is 23. The method listed is ASTM D790.
- The Ross flex test result is more than 100,000. The method listed is ASTM D1052, reapproved 2019.
- Compression set is 7 percent. The method listed is ISO 815-1 and ASTM D395, type B, 72 hours at 23 °C.
Three notes help when reading those values.
First, check the method column. Tear resistance, tensile strength, and elongation at break all list ASTM D638. Shore A hardness lists ASTM D790. Ask Stratasys to confirm the test method behind each value before it goes into a specification, a qualification file, or a customer document.
Second, the compression set was measured at 23 °C. The sheet reports no compression set at an elevated temperature. A seal or a gasket carries its load at its working temperature, so that question stays open.
Third, the sheet does not say how the test specimens were printed or oriented, or how many were tested. Ask whether the values change with build direction.
Biocompatibility
The sheet reports three biocompatibility results, and all three read Passed.
- Cytotoxicity, tested to ISO 10993-5 and USP 87.
- Sensitization, tested to ISO 10993-10.
- Irritation and intracutaneous reactivity, tested to ISO 10993-23.
An asterisk on each result points to a footnote that says the results are according to the supplier procedure. Each row also names a steam-sterilized test article.
Other statements on the sheet read more broadly than the table. The opening text describes ISO 10993 biocompatibility certification. The feature list says biocompatibility holds without the need for additional treatment, and the text says the material is biocompatible after the normal postprocess, with no extra steps or specific solvents. The table, though, names steam-sterilized test articles. The sheet does not say whether steam sterilization is part of the normal postprocess. It does not say who ran the tests. It reports three results and no other ISO 10993 tests.
Questions to ask Stratasys before a skin-contact part depends on this:
- What was the test article, and how was it postprocessed?
- Who ran the tests, and can the reports be seen?
- Do the geometry, wall thickness, and postprocess of the planned part match the test article?
The aging chart
The Thermal Properties section holds one chart. It shows an accelerated aging test at 80 °C. The chart plots tear strength, tensile strain, tensile strength, and Young's modulus at 1, 3, and 5 simulated years, as a percent change from the starting value. It has no data labels. The values below are read from bar heights, so treat them as approximate.
- Tear strength rises, by about 15 percent at one simulated year and about 45 percent at three and five years.
- Tensile strain falls to about 70 to 80 percent of its starting value.
- Tensile strength changes little, within about 10 percent above the start.
- Young's modulus rises, by about 55 to 85 percent.
Together, those bars show a material that gets stiffer and stretches less as it ages, while tear strength goes up. For a part whose job is to stay soft, such as a seal or a band, the modulus change is the number to look at first.
The sheet does not say how 80 °C converts to simulated years. It does not say how many specimens were tested or what the error bars show. The section is titled Thermal Properties, yet the sheet gives no continuous-use temperature range and no low-temperature behavior.
Printable features
Stratasys lists the smallest features it evaluated on a test part.
- The maximum unsupported overhang length is 3 mm.
- The maximum span length is 7 mm.
- The minimum unsupported overhang angle is 15 degrees. The sheet does not say which reference the angle is measured from.
- The minimum vertical wire diameter is 0.5 mm at 1 mm height, and 1 mm at 3 mm and 5 mm height.
- The minimum unsupported wall thickness is 1 mm at both 5 mm and 10 mm height.
- The minimum hole diameter is 2 mm in the Z direction and 2 mm in the XY plane.
These are evaluated values from one test part. They do not guarantee the same result on a different geometry. A vent or a drain hole under 2 mm, or an unsupported wall taller than 10 mm, falls outside what the sheet reports. A test print of the real part answers the question the table cannot.
What the sheet leaves out
The Chemical Properties section holds the biocompatibility results, a cured density of 1.3 g/cm³, and a 24-hour water absorption of 2.1 percent. Chemical resistance data is not on the sheet. Nor are price, shelf life, storage conditions, or the postprocess steps in detail. The sheet does not compare the material with a molded silicone grade. Some of that may sit in other Stratasys documents. This sheet does not say.
What to take from it
- Read the sheet as a vendor's statement of its own tests, not as independent data.
- Ask for the test method behind each value before citing it.
- Confirm the biocompatibility test article and its postprocess against the planned part.
- Ask for the aging data as numbers, with the conversion from 80 °C to simulated years.
- Check the evaluated features against the real geometry, then print a test part.
- Confirm access to an Origin system, since the material was developed exclusively for it.
References
Stratasys. P3 MED Silicone 25A by Shin-Etsu, Material Data Sheet, P3 DLP. Document MDS_P3_MEDSilicone25A_0726b. https://www.stratasys.com/siteassets/materials/materials-catalog/p3-materials/med-silicone-25a/mds_p3_medsilicone25a_0726b.pdf